Device Manager

Key Features

  • The full microscope configuration is displayed visually and controlled by the user using a mouse.

  • Simplified work with multi-device hardware configurations. Double-clicking on a device in the scheme opens its control pad.

  • Multiple hardware configurations (multiple independent microscope setups) can be set and easily switched between them.

  • Lightpath concept - optical paths are colored to match the defined wavelengths. The program watches the current hardware state and produces a warning if image acquisition will not be possible. See Lightpath tool bar.

  • Simplified definition of Optical Configurations – only devices enabled in the current hardware configuration are visible within the View > Acquisition Controls > OC Panel panel.

Hierarchy of Configurations

Hardware Setup

Sets which microscope is used for the ongoing hardware configuration.

Hardware Configuration

Determines which devices are used together, other devices can be turned off.

By default, the Acquisition hardware configuration is created. The user then adds the connected Cameras and Devices using the buttons. Lightpath combinations are automatically calculated in the Lightpath Selection table. The user usually does not need all the devices connected so any camera/device can be turned off creating a new device combination. This new hardware configuration can be saved by clicking on the Create New Hardware Configuration button and a new tab with a given name is created. The user can switch to a different configuration using the tab and clicking Activate, using the Devices > Select HW Configuration command or on NIS-Elements startup. Devices which are disabled for the selected configuration are not loaded and can be switched off to make the startup faster.

Lightpath

Groups devices (from the hardware configuration) used in one observation mode (EPI, DIA, TIRF, confocal, etc.).

The table inside each hardware configuration contains Light Paths which represent all available light paths (light tracks between the cameras and illumination devices present in the current configuration). Simply stated a light path tells which devices are used at the same time. Click on the name of a light path to highlight it in the image. Check Edit to select which light paths will be shown in the current hardware configuration.

Different light paths are usually saved under different hardware configurations to make the microscope operation most powerful and efficient.

Note

Hardware buttons of the Nikon Ti2-E microscope can be set to switch to a predefined light path. In the Devices > Device Manager , select Configure... in the context menu over the microscope, switch to the Programmable Buttons tab, choose a button and click on its drop-down arrow, click Select Lightpath and set a light path from the list.

Optical Configuration

specifies the settings of each device in a light path. Each optical configuration is associated with one light path.

Startup Hardware Configuration Selection

If you are running NIS-Elements for the first time, you will be asked to select a microscope and define the first hardware configuration. Click Select Microscope to open the New Setup window. Select the microscope connected to your computer and click Create Setup. The Device Manager window opens. Use this manager to add all cameras and devices connected to the microscope, then create hardware configurations and lightpaths.

You will be asked to select a hardware configuration every time you launch NIS-Elements (unless you selected the Select automatically on startup option). Only the cameras and devices present in the selected configuration will be connected. You can switch to another hardware configuration later in the device manager or using the Devices > Select HW Configuration command. If you need to create a new hardware configuration without loading an existing one, please select the predefined configuration called Offline Analysis during and then create a new configuration in the Devices > Device Manager .

Device Manager Window

  1. Hardware Setup tabs - select which microscope is used. More setups are added by the button.

  2. Hardware Configuration tabs - select a hardware configuration. More configurations are added by the button.

  3. Lightpath Selection table - displays all available light paths for the current hardware configuration and selects the active light path. Use Edit to select which light paths will be hidden/shown and change their order using the arrows. To add/remove the light path with all devices, click Settings... and check/uncheck All Devices.

  4. Microscope/Cameras/Devices list - shows all connected microscopes, cameras and devices. To add more, use the button.

  5. Microscope scheme - shows a visual microscope scheme of the current setup. The system can be configured directly from this scheme using mouse dragging and context menus.

  6. Device Manager Settings - Export... exports the selected hardware setups (hardware configurations, light paths, optical configurations, etc.) into a .dat file whereas Import Setups... can be used to load the previously exported setup. Help opens this help page.

Adding Cameras and Devices to Hardware Configuration

  1. Display the device manager by the Devices > Device Manager command.

  2. Select and activate a hardware setup and a hardware configuration in the top tool bar. In case you run the device manager for the first time, you will be prompt to select a microscope from the list of installed drivers. This will create a hardware setup and hardware configuration.

  3. Click the Cameras caption to add a camera, click Devices to add a device such as an illuminator, a motorized stage, etc.

  4. Select the model of your device from the list and click Add ... button.

    Note

    Only cameras and devices whose driver was selected during the installation process are listed.

  5. A drawing of the camera or device appears in the window either mounted to a default position or floating next to the microscope. Drag it by mouse so that the visual position matches the physical setup. Available positions are highlighted yellow.

    Note

    Non-standard mounts which are not recognized by the software can be connected to the microscope by adding the Coupler Universal device which can connect any camera or device to any port. The Coupler Universal Switcher can connect two cameras or devices to a single port.

  6. Right-click the camera or device to display a context menu where various actions may be made.

Connecting a Device to NIS-Elements

Before you get to work with NIS-Elements, all hardware accessories should be connected properly to the system. In most cases, the following basic procedure is sufficient to connect a device successfully:

  1. Install NIS-Elements, and select the appropriate device(s) during the installation.

  2. Connect the device to the PC and switch the device ON.

  3. Run NIS-Elements, and run the Devices > Device Manager command to open the Device Manager window.

  4. Select a microscope hardware setup or create a new one.

  5. Click the Add button next to the Devices caption to add the device to NIS-Elements.

  6. Select the device from the list of installed devices and click Add Device.

  7. Click on the device in the scheme and drag it to the highlighted hotspot so that the connection matches your physical microscope setup.

  8. Create a hardware configuration with light paths suitable for the experiment.

  9. Close the Device Manager.

If a peripheral device is connected to NIS-Elements, some extra commands and features become available. This concerns typically motorized XY stages and Z drives, but also other devices.

Note

MULTIZOOM AZ100M, ECLIPSE LV series, ECLIPSE MA200, or ECLIPSE L200N/300N microscopes can not be connected from NIS-Elements while the setup tool of each microscope is active. To connect any of these microscopes to NIS-Elements, exit the setup tool and then run NIS-Elements.

Renaming of devices

Either logical or physical device can be renamed by the user. Just right-click the device name and select the Rename... command from the context menu. The user-defined names can be handy in two cases:

  • You are using two logical devices having matching names, but you need to uniquely identify them (e.g. from a macro).

  • You are used to call a device with another name and would like to rename it in order not to be confused by the predefined name any more.

Lightpath Scheme

This panel invoked by the Devices > Show Lightpath Scheme command displays the active hardware configuration. Physical tracks of the light - so called lightpaths - are indicated in the scheme by colorized lines. If the light path is blocked (dashed line), a warning symbol () appears by the device which prevents the light from passing through. The same warning symbol displayed in the list of lightpaths indicates that some device in that particular light path are not set correctly. Such lightpaths will not be saved. Thanks to this indication the user can quickly detect possible errors in the setup (e.g. wrong filter cube assignment).

Microscope buttons, shutters, lights, etc. can be controlled directly by clicking on them in the scheme. Double-clicking on a device opens its control pad. Devices not participating in the current light path are outlined in gray and can still be controlled.

Note

Stimulation light paths are not shown in the Lightpath Scheme. If you need to use the stimulation light path for acquisition and not only for stimulation, enable the acquisition light path under the Edit button in the Devices > Device Manager window. If this light path is missing, make it visible by clicking the Settings... button and checking it.

Context menu over a device

Activate Lightpath with this device

Finds and activates a light path being the most similar to the currently selected light path and containing the device being clicked at the same time.

Rules for "Lightpath name" Lightpath

Specifies further rules for the clicked device in the current active light path.

Set State via OC, Set Position via OC

Controls the device state/position/settings using Optical Configurations when the current light path is selected.

Ignore

Does not change the device state/position/settings when the current light path is selected.

Active Shutter, Active Stimulation Shutter

A shutter can be used as active (preferred) in the current light path and all other shutters will be set to their default behavior.

Position

Always keeps the device in the position selected from the drop-down menu when the current light path is selected.

Configure...

Opens the configuration dialog of the clicked device.

Lightpath Scheme Settings

In this window, you can change the display colors of the light paths, display or hide the list of light paths within the panel, display or hide text labels over the devices and turn on/of the warning message when a light path is changed and positions of non-motorized devices do not match.

Lightpath Colors

Pseudo Color (Red/Green/Blue)

Excitation is red, emission is green, stimulation is blue.

Real Color - Mixed Color

One color is used calculated from all present wavelengths.

Real Color - Multiple Colors

Each wavelength will be displayed as a separate line.

Fit to Screen

Adjusts the zoom so that the microscope scheme fits into the Lightpath Scheme pad. The scheme can be zoomed in/out using the mouse wheel.

Device Manager...

Opens the Devices > Device Manager window.

Dual, Triple, Quad Camera Setup

(requires: Dual Camera Support) (requires: Tripple/Quad camera support)

Some camera models may be used in a dual, triple or quad mode. The software combines their grayscale images into one multi-channel image.

  1. Run the Devices > Device Manager command.

  2. Add the Dualcamera device and two or more cameras to the configuration. Arrange them to match the physical setup.

  3. Right-click a camera and select the Make Dual Camera with command. A dashed line will connect the two selected cameras.

  4. Repeat the previous step for all of your cameras. For a triple camera setup use two Dualcamera devices connected in a row, for a quad setup use three.

See Dual, Triple, Quad Camera Setup for an example.

Introduction to Dual, Triple, Quad Camera Setup

Two, three, or four monochromatic cameras of the same type can be run in a multi-channel mode. In this mode, the application behaves as if a two, three, or four-channel camera was connected. The following cameras support this feature:

  • Andor Ixon (897 X3, 897 Ultra, Ixon 885)

  • Andor Neo

  • Andor Zyla (5.5 3-tap, 5.5 10-tap)

  • Hamamatsu Flash 4.0 (CL, USB3)

Dual Andor iXon Setup Procedure

  1. Check the Wiring between Cameras

    The master camera Fire connector should be connected to the Ext. Trig. connector of the slave camera.

  2. Select Andor driver

    Run the Devices > Device Manager command.

  3. Add both of your Andor cameras.

  4. Add a Dualcamera device.

  5. Connect (drag) the Dualcamera device to the microscope port so that the dual camera setup matches your physical setup.

  6. Mount (drag) both of the cameras onto the Dualcamera device.

  7. Select one of the Andor cameras, reveal its context menu and select Make Dual Camera with... the second camera.

Dual Camera Settings

Master Camera, Slave Camera

Serial numbers of the connected cameras are listed in these pull-down menus. Select, which camera is which. The triggering cable is connected to the Fire connector on the master camera and, on the opposite end, to the Ext. Trig. connector on the slave camera.

Rotate 180°, Align, Flip, Flip Image of the Slave Camera

When optical paths of the two cameras differ in number of mirrors, you may encounter one of the images being flipped. Depending on the camera driver, these options are available for you to adjust orientation of the image channels so that they match.

Synchronization

Select the type of Synchronization to be used during Live and during No Delay acquisition.

Frame by frame

In this mode, it is guaranteed, that frames from both cameras were captured at the same time. On the other hand, lower frame rates can be achieved and some advanced settings get disabled.

On acquisition startup

In this mode, you can achieve very high frame rates and all settings are controllable. Even though single frames are not synchronized precisely, this mode is widely usable.

Enhance Timing Precision

Enhances timing performance of the cameras. The Arm and Ext. Trig connectors of the Master Camera must be connected by a cable. Then, exposure times on all cameras are controlled (triggered) by the master camera instead of being triggered from the software. This option is available only in the On acquisition startup synchronization mode.

Using a Multiview Device

A Multiview device (dual, triple or quad) enables you to divide the camera chip into halves/thirds/quarters and observe the chip parts in an overlay as a multichannel image.

  1. Run the Devices > Device Manager command.

  2. Add your mutli-view device to the configuration by the Devices button. If you cannot find your multi view brand in the list, use the generic MultiView device.

  3. Arrange the configuration to match the physical setup.

  4. Right-click the multi view device and select Configure Multiview, then select whether it is a Dual View, Triple View or Quad View.

  5. Click OK.

Note

  • Bypassing the multi view feature is done by switching to the light path with the MV(Out) postfix.

  • To run a multi view mode on a dual camera, connect two Multiview devices to one Dualcamera device, display a context menu over a camera and select Make Dual Camera with.

What are Logical Devices?

NIS-Elements handles hardware accessories using the concept of logical devices. There are features of different hardware devices which equal and therefore can be controlled equally. Such features are called logical devices. A typical logical device is Stage XY. Microscopes can be equipped with different XY motorized stages, although - regarding the user interface - they behave equally. One physical device (a piece of hardware) can contain one or more logical devices the list of which appear in the Device Manager after the connection is established.

Available Logical Devices

Analyzer

The analyzer is a polarizing filter placed in the optical path between the specimen and the lamp. The logical device offers two states: ON (inserted) and OFF (extracted).

Aperture

This logical device is used for controlling apertures in the light path. It is used in complex microscopes rather than as a standalone device. Two parameters can be typically set for aperture devices, the state (on/off) and aperture size.

Condenser

A condenser is a two-lens combination located next to the illumination source in the optical path. Its purpose is to collect light and direct it to the specimen being examined. The corresponding logical device relates to a changer of different condensers.

Filter

This logical device controls filter changer movements. There can be several filter changers connected to NIS-Elements at a time. Each filter changer needs to be set up - filter types shall be assigned to positions of the changer:

  1. Display the filter changer control panel (Devices > Filters and Shutters or Devices > Microscope Control Pad ).

  2. Click the settings button , a window appears.

  3. Select one of the available positions which the filter will be assigned to.

  4. Click the ... button, a list of available filters appears.

  5. Select the filter name from the list and confirm it by OK.

  6. The filters can be moved within the already defined positions using the Up and Down arrow buttons.

Note

When browsing the list of filters, details about the currently selected filter are displayed on the right side of the window.

Illuminator

This logical device is used for controlling the specimen illumination remotely. There is no standard dialog box for the illuminator control. Each device handles this logical device via a user interface specially designed for it - typically containing one button for switching it on/off and a slider for regulating intensity.

Light Path

Some microscopes have more than one port where it is possible to attach a light source or a camera. This logical device can switch the illumination between these ports.

Microscope

This logical device is used to group standalone logical devices used in certain microscopes. To control the logical devices of a microscope from one control panel, select the Devices > Microscope Control Pad command.

Nosepiece

This logical device serves for controlling microscope objective changers.

See Assigning Objectives to Nosepiece Positions, Motorized, Intelligent and Manual Devices.

ND Filter

A neutral-density filter is a light absorbing filter whose absorption spectrum is moderately flat. It is used to reduce the illumination intensity within the optical path. The logical device offers two states: ON (inserted) and OFF (extracted).

PFS

Perfect Focus System - this logical device corresponds to the PFS physical device available with Nikon TE2000/TI microscopes.

Shutter

This logical device can control shutters installed in your system. This device is handled via the Devices > Filters and Shutters control panel or straight from the microscope control pad. You can select the type and rename the shutter by running a contextual menu command either within the Filters and Shutters control panel, Device Manager window, or in the main toolbar.

Note

There was a restricted number of shutters which could be connected to the system in previous releases of NIS-Elements and they were identified by Type (DIA, EPI, Aux1, etc...). The current version supports identification of shutters by custom Names, but the Type attribute can be still found in some windows (or used in some macro functions). This is to ensure backward compatibility.

Zoom

This logical device is used for controlling the zoom factor. Run the Devices > Zoom Configuration command to adjust the zoom settings.

Stage XY

Stage XY enables the movement of specimen within X and Y axes. The system offers user to control stage movements.

Stage Z

The Z Drive device enables movement throughout the Z axis direction.

TTL/Analog Input, Output

There are six more logical devices available when the NIDAQ controller set is installed:

TTL Input
TTL Output
Analog Input
Analog Output
Calibrated Analog Input and Calibrated Analog Output.

Each of these logical devices represents a connection to an external device which can receive (send) signals from (to) NIS-Elements. The number of available connections depends on the actual NIDAQ card.

See View > Acquisition Controls > ND Control Panel

Motorized, Intelligent and Manual Devices

For the purposes of this manual, we distinguish the following device types. Each type provides a different level of control from the software.

Manual

It cannot be controlled via the software neither the software is aware of its current state. For the purposes of image acquisition, the current state of such a device shall be set in the software manually so that the meta-data of the acquired image would be correct. This typically concerns a manual nosepiece or a filter changer.

Intelligent

The current state of the device is displayed within the software GUI, but the device cannot be controlled.

Motorized

The device can be fully controlled via the Microscope Control Pad or other panels.