Microscopes

Note

The complete list of cameras and devices supported by NIS-Elements is available in a separate document.

Nikon Ci-E Microscope

The CI-E Pad (fully armed) consists of the following portions:

Nosepiece

The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.

Condenser

You can insert or remove the condenser lens by these two radio buttons.

Note

Condensor setting (In/Out) can be bound to nosepiece positions when using the hardware remote control. In NIS-Elements, use optical configurations to achieve the same behavior (see Optical Configurations)

LED

The microscope LED illumination can be controlled from here.

Configure

A window with detailed information about the software opens when you press this button.

Nikon AZ100 Microscope

Note

You cannot use the handheld remote controller and foot switch while the MULTIZOOM AZ100M is connected to NIS-Elements.

The AZ100 Microscope Control Pad consists of the following portions:

Nosepiece

This pad indicates the nosepiece status: the currently selected objective and objectives magnification. The objective changer can not be controlled by NIS-Elements.

Use the button to select objectives from the database of objectives. See Assigning Objectives to Nosepiece Positions.

Zoom

Zoom magnification can be changed by moving the slider. The current zoom factor is indicated above. The user can calibrate the zoom manually via the button on the right. See AZ100 Zoom Calibration

Note

Since the zoom device is a part of the microscope equipment, concerning optical configurations, its settings are automatically treated as an inseparable part of microscope settings (there is no Zoom in the Used devices box).

Z Drive

The Z drive can be controlled via the microscope control pad in the following way:

  • Enter the exact position of the Z-drive. The range is <-80,0>

  • The current position can be memorized by pressing the Memory button. The position value will be remembered under the currently selected position.

  • To move the Z drive to a memorized position, select the desired value from the pull-down menu and press the Recall button.

Filters

The currently selected filter of the four available filter positions is indicated in this portion of the pad. The filter changer cannot be controlled by NIS-Elements. Filters assignment may be changed after pressing the button.

Condenser

These buttons indicate the current condenser lenses position. The condenser cannot be controlled.

AZ100 Zoom Calibration

The default calibration of the microscope zoom can modified by the following procedure:

  1. Run live camera image and place a calibration sample to the field of view (e.g. a ruler).

  2. Click the button in the Zoom section of the microscope pad. A dialog window appears.

  3. Change Zoom to the value you are about to correct/calibrate by the slider and click the Recalibrate Zoom button.

  4. Calibrate the live image manually. See also: Calibration.

  5. A new Corrected value is added to the Corrected Zoom Values table.

  6. Repeat this procedure for different zoom factors. The resulting calibration is a linear approximation of all the values in the table.

Dialog Options

Zoom

Changes zoom on the microscope.

Recalibrate Zoom Nx

Starts manual calibration for the current zoom. See also Calibration.

Reset

Resets the calibration table to default values.

Export, Import

The calibration table can be saved to a XML file and loaded later. Use these buttons and standard Open and Save dialog windows.

Use Software Backlash

Motorized devices achieve better positioning repeatability if a backlash compensation is used. The software backlash compensation ensures that the movement direction used to reach a certain position (of a zoom or a nosepiece) is always the same. For example, if you change zoom from 1x to 2x, the movement is direct, but if you change from 3x to 2x, the zoom changes to 1x first and than to 2x in order to reach the position from the same direction.

Nikon Eclipse LV Series Microscope

The LV (fully armed) Microscope Control Pad consists of the following portions:

Nosepiece

The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.

Caution

This behavior is valid if the Limit Control option is turned off in LV Setup. If Limit Control is turned on, the warning appears if a move is ordered from a lens with magnification 5x (or less) to a lens with Working Distance less then 1 mm.

Escape Z

Press this button to move the Z drive to the utmost position away from the sample. Use it when changing objectives or slides on the XY stage or during XY stage initialization. When the button is de-selected, the Z drive moves back to the most recent position.

Z Drive

The current Z position is displayed in the edit box. A particular value can be inserted as well and confirmed by pressing Enter to move the Z-drive.

Lower Limit

When using Z drive of the ECLIPSE LV series, make sure that the bottommost position of the Z drive (called Lower Limit) is set to avoid collision between the specimen and the objective.

Set Limit

Click this button to set the current Z position as the Lower Limit.

Clear Limit

Click this button to discard the Lower Limit setting.

Please see for details.

Lights

Intensity of episcopic and diascopic illumination sources can be controlled from here. LED illuminators are also supported.

Aperture

75% sets the aperture stop to 75% of the objective pupil. The slider with the edit box set a custom aperture size.

Shutters

Controls the connected shutter.

Filter Block

There are 4 available positions in the changer. The first two have fixed filters for bright field (BF) and dark field (DF) microscopy. The other two are available for any filters.

Use the button to select the filters from the database of filters.

Observation

Defines the observation mode.

Nikon Eclipse L200N(D)/L300N(D) Microscope

The L200/L300 Microscope Control Pad consists of the following portions:

Nosepiece

The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.

Illuminator

A halogen lamp or the Nikon Intensilight illumination device can be attached to the EPI port of the microscope. Both illuminators can be controlled from here (see below). If two halogen lamps are attached, the illumination control works as the EPI/DIA switcher.

Aperture Stop

These two arrows control the EPI aperture stop equally to the buttons on the microscope body.

Note

It is not possible to save the aperture status to an optical configuration. The only way to adjust it is by these two buttons (or the buttons on the body).

Control

You can select from where the lamps will be controlled:

NIS-Elements

from the Control Pad

Microscope

from the microscope body

Nikon Eclipse MA200 Microscope

The MA200 Inverted Metallurgical Microscope Control Pad consists of the following portions:

Nosepiece

The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.

Zoom

Indicates the current state of the MA Zoom device. There are three predefined magnifications (1x, 1,5x, 2x).

Lights

A halogen lamp or the Nikon Intensilight illumination device can be attached to the microscope and controlled from here. If the halogen lamp is attached to the microscope, internal (manual) or external (via NIS-Elements) control can be selected on the back of the microscope body. LED illuminators are also supported.

Nikon MM 400/800 Microscopes

This measuring microscope connected to NIS-Elements provides two logical devices:

Nosepiece

Two types of nosepiece can be attached to the microscope, manual or intelligent.

See Motorized, Intelligent and Manual Devices.

After adding the MM microscope to the system via Device Manager, the Nosepiece logical device becomes available. Display its control panel by the View > Acquisition Controls > Nosepiece command. The type of the nosepiece must by defined via the hardware control pad in order to be recognized by NIS-Elements.

Z Drive

The Measuring Microscopes Z drive may be moved using the View > Acquisition Controls > XYZ Navigation control panel.

Nikon Ni-E Microscope

The NI-E Pad (fully armed) consists of the following portions:

Nosepiece

Control the objective changer by clicking on the buttons.

Use the button to select objectives from the database of objectives. See Assigning Objectives to Nosepiece Positions.

Zoom

An adjustable optical zoom is installed on the Rear microscope port. If the light path is switched to Rear a slider appears enabling to adjust the zoom factor.

Escape
Escape Z

Press this button to move the Z drive to the utmost position away from the sample. Use it when changing objectives or slides on the XY stage or during XY stage initialization. When the button is de-selected, the Z drive moves back to the most recent position.

Escape YZ

Press this button to move the Z drive to the lowest position and move the XY stage outward from the microscope body. This function is designed for easy changing of samples. When the button is turned off, the XY stage moves back to the original position but the Z drive stays at its lowest. It is because the user would have to re-focus anyway after a new specimen was inserted.

Clear

Turns off the Escape Z button but does not move the Z Drive anywhere and lets you refocus manually.

Lightpath

Select a port where the light will be directed.

Z Drive

Use the arrow buttons to move the Z drive by the specified distance. Or, enter the absolute distance to the Z edit box.

Lights

Control panels for controlling DIA and Intensilight (EPI) illumination appear if the corresponding devices are connected.

Shutters

Buttons in this section control all connected shutters.

Filters

Control the available filter changers by clicking the buttons.

Use the button to select filters from the database of filters.

Aperture, Field Stop, ND Filter

Control these aperture-like devices by moving the slider or entering numbers to the nearby fields.

Condenser

Control the condenser changer by selecting a position from the pull-down menu.

Use the button to select condenser from the database of filters.

Sleep

This button turns off some of the motorized microscope parts in order to minimize electromagnetic radiation of the device which might affect the specimen.

Recipe

Check this item to use Ni recipe for capturing ND experiments. Define which port is used for the recipe.

Configure

A configuration window appears where it is possible to specify microscope options.

XY Software Limits, Z Software Limits

Limits to XY stage and Z drive movements can be specified. Insert maximum and minimum distances for each axis. Just the maximum/minimum of the Z drive can be specified depending on its type (whether it moves the stage or the nosepiece). See for further details.

Speed

Select speed for XY and Z motors.

Accuracy

Accuracy can be set to XYZ stages equipped with encoder.

Keep Z position

This function guards the stage position in between stage movements so that the stage stays perfectly in its last position thus improving the overall experiment accuracy.

Filter Turret Speed

Speed of the filter turrets can be reduced by selecting the Slow option.

Shutter ports

Specify what type of shutters are connected to the ports on the back of the microscope body (none, EPI, DIA, AUX).

Show Sleep

This option enables you to hide or show the Sleep button on the Ni-E Pad.

Show Microscope Settings

Displays the Current Settings to Microscope Mode at the bottom of the Ni-E Pad. Select to which microscope mode will be the current settings assigned. Editable name of the selected microscope mode is displayed next to this pull down menu. You can enter arbitrary four characters as a name. If you press the Assign to Microscope button, current settings are assigned to defined microscope mode.

Import/Export Settings

Saves the configuration settings into a .txt file using Export... and imports the dialog settings from a .txt file using Import.... During the import, it is possible to select which parameters are imported using the check boxes.

Nikon Ni-U/Ni-L Microscope

The Ni-U and Ni-L microscopes are simplified versions of the advanced Ni-E microscope. Please see Nikon Ni-E Microscope for the description.

Nikon SMZ Stereomicroscopes

Note

Supported models are: 18, 25, 754T, 800, 800N, 1000, 1270, 1270i, 1500

Depending on the type and setup of your SMZ, the control pad can contain different options.

SMZ Pad Options

Nosepiece

An intelligent nosepiece which indicates the current light-path setting (Stereo/Mono).

See Motorized, Intelligent and Manual Devices.

Escape

Click the Escape Z button to move the Z Drive up by the distance specified in the rightmost edit box. Click it again to move the Z drive back to its original position.

The Cancel button resets the escape status (โ€œescaped or not escapedโ€) to โ€œnot escapedโ€. E.g. If you pressed the sequence of buttons: Escape Z - Cancel - Escape Z, the Z drive would move twice the defined distance up.

Zoom

This control enables you to specify the zoom value.

ALZ

ALZ (Auto Link Zoom, supported by Nikon SMZ 25 only) can be activated to keep the magnification constant (same FOV) even if the objective lenses are switched: Objective magnification x Zoom = constant.

Zoom slider

Zoom slider defines the zoom of the microscope (supported by SMZ 25 only). Use the edit box next to the slider to enter a precise zoom value.

Memory

Memory assigns the current Zoom/Z setting to the position selected in the pull-down menu. To recall a setting, turn on Recall and select the setting from the pull-down menu.

Recalibrate

The Recalibrate button opens the Manual calibration window used for calibrating the current zoom magnification.

Note

For a correct zoom calibration on multiple SMZ objectives, follow this procedure:

  1. Select the first objective on your nosepiece.

  2. Click Recalibrate... and follow the calibration instructions (see: Objective Calibration).

  3. Switch to the next objective, go to Calibration > Objectives and Recalibrate it.

  4. If your microscope has more than two objectives, repeat step 3.

Z Drive

Standard Z Drive control. The only difference from other microscopes is that the units used are millimeters.

Note

Because the minimum step of the SMZ Z drive is 1.6 ยตm, it is not possible to move exactly by the amounts displayed in the XYZ Navigation (View > Acquisition Controls > XYZ Navigation ) and by values lower than this step. XYZ navigation is a generic GUI and its values are not changing via Z drive, so SMZ moves to the next step with the best precision which it can. Use 1 and 0.5 up arrow buttons to get over your desired Z value and then precise the number by 0.05 and 0.003 down arrows.

Lights

Control panels for controlling DIA and Intensilight (EPI) illumination appear if the corresponding devices are connected.

Shutters

Buttons in this section control all connected shutters.

Filters

The microscope may be equipped with a motorized (or intelligent) filter turret.

Configure

A configuration window appears where it is possible to specify microscope options.

Z Software Limit: [mm]

The bottom limit to the Z drive movement can be specified. Click the Z button to insert the current position as the limit. See for further details.

Speed

Select speed for the Z drive motor.

Using Nikon SMZ in the Manual Mode

The Nikon SMZ product line is very variable, some models are fully manual, some contains intelligent or motorized parts. If you plan to use your SMZ in the manual mode, connect the microscope as Nikon SMZ Manual in the device manager.

See Motorized, Intelligent and Manual Devices.

Pad Options

Nosepiece

Specify the objectives mounted to your nosepiece via the button.

Zoom

Zoom factors selected for display in the configuration window.

Recalibrate Zoom

You can modify calibration for the selected Zoom. Click the button and proceed to manual calibration.

Filters

If a turret is selected in the configuration window, you can select the used filters via the button.

Configure

Click this button to specify properties of the SMZ microscope used.

Model

Microscope model. Other options in this window depends on the selection.

Nosepiece positions

Select how many objective positions will be displayed in the pad.

Turret mounted

Is the turret mounted? Select this option.

Coaxial EPI

Having this accessory mounted on the microscope changes the zoom factor. Select it if you have it.

Calibrated Zooms

You can modify the list of displayed zoom buttons by the Add and Delete buttons. The Reset button loads the default list of zooms and their calibrations.

Nikon Ti2-E Microscope

Parts of the Ti2-E Pad

The Ti2 (fully armed) Microscope Control Pad consists of the following portions:

Nosepiece

Control the objective changer by clicking on the buttons.

Use the button to select objectives from the database of objectives. See Assigning Objectives to Nosepiece Positions.

Note

If the installed nosepiece supports the correction collar, the position number 5 gets disabled.

Escape Z

Press this button to move the Z drive to the utmost position away from the sample. Use it when changing objectives or slides on the XY stage or during XY stage initialization. When the button is de-selected, the Z drive moves back to the most recent position.

The Cancel button is displayed in the escape mode and can be used to clear the escape flag - i.e. stay at the current Z escape position and unlock the Z control.

Switchers

Camera / EYE notifies about the light path direction if an eyepiece having two paths (eyepiece and camera) is connected.

Lightpath

The light path control buttons correspond to the buttons placed on the microscope body.

Note

Bertrand lens is automatically removed from the light path when switching from EYE to Camera.

Z Drive

Movements of the integrated Z drive are performed by the up/down arrows. Accuracy can be set in the pull-down menu on the right. The Accuracy setting affects the Z drive speed slightly.

Correction Collar

This section controls the objective correction collar. The slider and the edit box defines the correction collar position.

Note

If the correction collar is used with a confocal microscope, the recommended pinhole size is 1.2 AU.

Click this button to open the ACC Calibration dialog window. See Automatic Correction Collar Adjustment.

Slider

Use the slider to set the correction collar value. Hover the cursor over the slider to display a tooltip table with the cover glass thickness and recommended correction collar values for particular temperatures.

Named presets of the correction collar values to the pull-down menu on the left of this button. Here you can add values to the list, name them and sort them.

WID (Water Immersion Dispenser)

If the dispenser device is installed correctly and a water immersion objective is activated in the microscope pad, these three buttons can be used:

Initi

Starts supplying water to the tubes and on the objective.

Replenish

Water evaporates in time. Once initiated, automatic replenishment is started and the period is set to 30 minutes. Shall the water be replenished earlier, use this button.

Discharge

Empties the tubing. Use this when you have finished the experiment.

See also Using Water Immersion Dispenser.

PFS

This optional accessory can be controlled via the pad.

PFS

The main PFS button turns the perfect focus system on and applies the current PFS offset. If the current objective supports two materials of the sample carrier (glass slide, plastic dish), a pull-down menu appears next to this button where the actual material in the sample holder shall be selected.

Dichroic

This button moves the motorized dichroic mirror for the PFS in and out. PFS works only if set to the In position.

Offset

If the PFS is on, the offset value can be typed right in the edit box.

Find PFS Offset

Turns the PFS on and finds the offset value so that it corresponds to the current Z-drive position.

Lights

The light source can be turned on/off using the button. Use the slider to set the light intensity or enter an exact value (percents or voltage) into the edit box.

In case of LED DIA Illumination the DIA LED button is added to the top toolbar so that it can be used in Optical Configurations as an active shutter.

Shutters

All available Ti2 shutters are controlled from here. When Nikon Intensilight is connected, INTS shutter appears. FL-Up and FL-Lo shutters correspond with the shutters in the filter turret.

Filters

Control the available filter changers by clicking the buttons.

Use the button to select filters from the database of filters.

Condenser

Control the condenser changer by selecting a position from the pull-down menu.

Use the button to select condenser from the database of filters.

External Phase

Can be used to select the External Phase Ring for the eyepiece.

Bioluminescence Mode

This mode is used to prevent any potentially harmful effects caused by the electronics inside the microscope which may be potentially harmful to the specimen or otherwise interfere with the acquisition process. Use the button to temporarily turn the electronics off and back on.

Display or hide this option in the General tab of the microscope configuration dialog.

Configure

Opens the configuration window.

General tab

Note

If a halogen lamp is connected to the system, โ€œMount Halogen Lampโ€ option has to be checked. This lamp requires an amount of time to be fully turned on and off. You can adjust the On/Off Time [msec] in the edit boxes which delays all experiments using the halogen lamp on/off command.

Show Bioluminescence Mode On Ti2 Pad

Does as it says.

Display LED

Turns the front panel LED diode ON.

Allow Parfocal Correction

โ€œAllowTi2ParfocalCorrectionโ€ has to be enabled first in the SetupOptions.xml file in the NIS-Elements installation folder.

Output Port Options

Set the ratio for dividing light between ports when the Right port is selected.

Import/Export Settings

Saves the configuration settings into a .txt file using Export... and imports the dialog settings from a .txt file using Import.... During the import, it is possible to select which parameters are imported using the check boxes.

XY tab

Sets the software limits for the XY, speed and accuracy.

Z tab

Sets the software limit, speed, Open/Closed Loop Control, accuracy and Keep Z position.

I/O tab

The main box of the Ti2-E has two connectors which can be used to establish a connection using 8 BNC I/O with additional 16 BNC with the Extension box. Select the usage of the BNC connectors in the pull-down menus:

TTL In

Enables the NIDAQ resource and controls accessories in the Triggered Acquisition and Illumination Sequence windows. PFI is selected per camera.

Camera Exposure In

FIRE signal from the camera is connected to this pin. It enables the user to use the Ti Recipe (recipe is turned on).

Camera Exposure Out

copies the camera out signal (works like a hardware โ€œT connectorโ€).

Sensor

Output signal: Hi = something is moving, Lo = nothing is moving.

PFS Status (Focus)

Hi = in focus, Lo = other states

PFS Status (Vicinity)

Hi = in range but not in focus, Lo = other states

PFS Status (Up/Down)

Hi = in focus or in range and over the sample, Lo = other states

PFS Status (Surface)

Hi = in range, Lo = out of range

TTL Output

If set, Output Signal Toggle on extended IO %d (two state button) or Output Signal High on extended IO %d Firmware Function in the Programmable Buttons tab can be assigned to a joystick button or microscope body button.

TTL Output Pulse

If set, Output Signal 5V 10ms on IO Firmware Function in the Programmable Buttons tab can be assigned to a joystick button or microscope body button.

Trigger Camera

Select this option to send the signal to the โ€œtrigger inโ€ connector of a camera during a continuous scan. A continuous scan is used for example for fast detection of well plates used by the Sample Navigation panel.

Programmable Buttons tab

Defines the programmable buttons on the microscope body and joystick. Each button can be used to run a specific macro (Run Macro), to run a command (Command List), to capture (Capture button), to switch to a selected light path (Select Lightpath) or to run a Firmware Function.

Note

When NIS-Elements is running, do not assign any Firmware Functions from outside tools and applications. If you still need to assign functions from external tools and applications, close NIS-Elements first.

Macro commands work with the programmable buttons only during the NIS-Elements session and override the firmware functions. When NIS-Elements is closed, all programmable button assignments are restored to the firmware state.

LAPP tab

Mounts the LAPP system. If a motorized or intelligent branch is mounted, it is automatically added and cannot be removed.

See: View > Acquisition Controls > LAPP Pad, Motorized, Intelligent and Manual Devices.

Filters

Two motorized filter turrets can be installed on the microscope. Their position (Upper Layer or Lower Layer) is detected automatically. In some specific cases, the positions could end up switched. In this tab, you can select the position of each turret manually.

Keep Z position

This function guards the stage position in between stage movements so that the stage stays perfectly in its last position thus improving the overall experiment accuracy.

The function cannot be used and is automatically turned off when using PFS.

Note

All the commands used in Ti pad for moving Ti accessories are recorded in the Command History (Macro > Command History).

All the microscope settings done inside NIS-Elements or in the Ti2Control App are stored in the microscope from which they are loaded on startup.

Guide

Runs the Ti2 Control where it is possible to set microscope accessories and options and execute โ€œAssist Guideโ€ suitable for making preparations of microscope observations. See the โ€œTi2 Control manualโ€ book for more information.

Automatic Correction Collar Adjustment

  1. Place the fluorescent beads sample (sample where PSF can be obtained) onto the stage. Or else, place the beads sample or biological specimen on the stage. In case of beads, find the less densely populated region where the number of beads within the FOV is about 10.

  2. Click on the Live button and enable Keep Auto Scale in the Live window.

  3. Display the Histogram (View > Visualization Controls > Histogram ) and adjust light and exposure time so that no saturated pixels are present in the image. After the adjustment, Freeze the live image.

  4. Click on the button in Ti2 Pad to show the ACC Calibration dialog window.

  5. Select the Sample Type in this window. Click on Define ROI to show the Live window.

  6. Move the motorized stage to place the target image (isolated beads or cells of interest) in the center of the FOV and bring it into focus. Set ROI to the target image in the Live window.

  7. Click on Confirm ROI in the ACC Calibration window to close the Live window.

  8. Sample Z Range and Search Range should be set as follows based on the type of objective.

    Table 2. 

    Objective/Sample TypeSample Z RangeSearch Range
    Range*Z Step*MinMaxStep Count
    100x Oil, Beads5.00.20-4960132807
    100x Oil, Biological Specimen2.50.05-4960132807
    60x Water, Beads30.00.20-624062407
    60x Water, Biological Specimen15.00.05-624062407

  9. Click Run in the ACC Calibration window to run auto correction. Z-Stack image acquisition is run and correction collar is adjusted automatically. When the adjustment is finished successfully, the correction collar is moved to the adjustment position.

Note

Approximately a two hour warm-up period is required after switching the microscope on.

For the correction ring adjustment, images acquired by the Epi-FL illuminator unit can be used as well as N-SIM images.

Hardware restrictions 
Please use one of the following cameras: Andor iXon3, Andor iXon Ultra or Hamamatsu ORCA Flash4.0. For the Z-Stage, please use MCL Piezo Z. If ACC is performed with a camera or a Z-stage different than those mentioned, correct results cannot be obtained.

Correction Ring Manual Adjustment

  1. Place the fluorescent beads sample (sample where PSF can be obtained) onto the stage.

  2. Find the less densely populated region where the number of beads within the FOV is about 10.

  3. Click on the Live button and enable Keep Auto Scale in the Live window.

  4. Display the Histogram (View > Visualization Controls > Histogram ) and adjust light and exposure time so that no saturated pixels are present in the image. After the adjustment, Freeze the live image.

  5. Move the motorized stage to put the isolated fluorescent bead image in the center of the FOV and bring it into focus.

    Tip

    Vertical movement can be adjusted by the mouse wheel after clicking on the live display.

  6. Open View > Acquisition Controls > ND Acquisition .

  7. Click on the Z tab. Press Symmetric mode defined by range and then Home to set the current focus position to the center of the stack.

  8. Set the Z stack so that the Z-Step is 0.12~0.16 ฮผm with 25~30 Steps.

    Click Run now to start image acquisition.

  9. On the acquired image window, click on the Show Slices View icon to show the cross-sectional images of XZ and YZ.

  10. From the Mode pull-down menu select Slices.

  11. Set the display magnification to 200%.

    Open View > Visualization Controls > LUTs and lower the upper limit of LUTs to make it easier to see the defocus spread.

  12. Move the correction ring by one position to refocus.

    Press Home of the Z series tab in the ND Acquisition (About ND Acquisition) to set it to the center position of the Z stack.

    Click Run now to acquire Z stack images in the same way.

    When the capturing is done, move the position of the correction ring and repeat the same process.

  13. By changing the position of the correction ring, compare the series of Z stacks and select the position so that the symmetry in the Z direction is the best. Adjustment of the correction ring is completed.

    In the examples below, the symmetry is best at 0.18.

Note

Approximately a two hour warm-up period is required after switching the microscope on.

For the correction ring adjustment, images acquired by the Epi-FL illuminator unit can be used as well as N-SIM images.

Using Water Immersion Dispenser

  1. Suppose the device is connected properly, a Water Immersion Dispenser section appears int the microscope pad. An immersion objective must be selected in the nosepiece to have all the buttons enabled.

  2. Click the Init button to add the first water on the objective. A progress dialog appears.

    Note

    If the objective is in focus position, it will move to the escape position for the time needed for initialization.

  3. Information about the time to the next automatic replenishment starts appearing below the buttons. E.g.: Next Replenishment in 30 min. If you need to replenish the water sooner than indicated, click the Replenish button.

  4. After the experiment is finished, discharge water from the tubes by the Discharge. This is important in order to keep the tubing clean. If you forget, the software will prompt you to discharge the dispenser upon exit.

  5. At last, do not forget to dry up the objective and the slide so that these parts stay clean too.

Tip

Behavior of the device can be customized in Windows registry on the following path:

HKEY_LOCAL_MACHINE\SOFTWARE\Laboratory Imaging\Misc\Ti2\WID

Settings for each objective can be edited as well (there are sub-folders named with objective codes).

Note

All Water Immersion Dispenser events are logged per microscope connection in a separate file into the following location:

C:\ProgramData\Laboratory Imaging\Platform\Logfiles\WIDRecord_*.log

where * is the date.

Nikon Ti2-A Microscope

Most features of the Ti2-A Microscope are similar to the Ti2-E Microscope. Main difference is that Ti2-A microscope uses only intelligent models of Shutters, Filters and Condenser.

See Motorized, Intelligent and Manual Devices.

Please see Nikon Ti2-E Microscope for full description of the microscope features.

Nikon Ti2-U Microscope

The manual version of the Ti2 inverted microscope. The microscope is not connected to the PC, correct microscope settings should be selected by the user (objective, filter, zoom) in order for the resulting images to contain correct metadata (calibration, etc.).