- Nikon Ci-E Microscope
- Nikon AZ100 Microscope
- Nikon Eclipse LV Series Microscope
- Nikon Eclipse L200N(D)/L300N(D) Microscope
- Nikon Eclipse MA200 Microscope
- Nikon MM 400/800 Microscopes
- Nikon Ni-E Microscope
- Nikon Ni-U/Ni-L Microscope
- Nikon SMZ Stereomicroscopes
- Nikon Ti2-E Microscope
- Nikon Ti2-A Microscope
- Nikon Ti2-U Microscope
Note
The complete list of cameras and devices supported by NIS-Elements is available in a separate document.
The CI-E Pad (fully armed) consists of the following portions:
The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.
You can insert or remove the condenser lens by these two radio buttons.
Note
Condensor setting (In/Out) can be bound to nosepiece positions when using the hardware remote control. In NIS-Elements, use optical configurations to achieve the same behavior (see Optical Configurations)
The microscope LED illumination can be controlled from here.

A window with detailed information about the software opens when you press this button.
Note
You cannot use the handheld remote controller and foot switch while the MULTIZOOM AZ100M is connected to NIS-Elements.
The AZ100 Microscope Control Pad consists of the following portions:
Enter the exact position of the Z-drive. The range is <-80,0>
The current position can be memorized by pressing the Memory button. The position value will be remembered under the currently selected position.
To move the Z drive to a memorized position, select the desired value from the pull-down menu and press the Recall button.
This pad indicates the nosepiece status: the currently selected objective and objectives magnification. The objective changer can not be controlled by NIS-Elements.
Use the button to select objectives from the database of objectives. See Assigning Objectives to Nosepiece Positions.
Zoom magnification can be changed by moving the slider. The current zoom factor is indicated above. The user can calibrate the zoom manually via the button on the right. See AZ100 Zoom Calibration
Note
Since the zoom device is a part of the microscope equipment, concerning optical configurations, its settings are automatically treated as an inseparable part of microscope settings (there is no Zoom in the Used devices box).
The Z drive can be controlled via the microscope control pad in the following way:
The currently selected filter of the four available filter positions is indicated in this portion of the pad. The filter changer cannot be controlled by NIS-Elements. Filters assignment may be changed after pressing the button.
These buttons indicate the current condenser lenses position. The condenser cannot be controlled.
The default calibration of the microscope zoom can modified by the following procedure:
Run live camera image and place a calibration sample to the field of view (e.g. a ruler).
Click the
button in the Zoom section of the microscope pad. A dialog window appears.
Change Zoom to the value you are about to correct/calibrate by the slider and click the button.
Calibrate the live image manually. See also: Calibration.
A new Corrected value is added to the Corrected Zoom Values table.
Repeat this procedure for different zoom factors. The resulting calibration is a linear approximation of all the values in the table.
Dialog Options
Changes zoom on the microscope.
Starts manual calibration for the current zoom. See also Calibration.
Resets the calibration table to default values.
The calibration table can be saved to a XML file and loaded later. Use these buttons and standard Open and Save dialog windows.
Motorized devices achieve better positioning repeatability if a backlash compensation is used. The software backlash compensation ensures that the movement direction used to reach a certain position (of a zoom or a nosepiece) is always the same. For example, if you change zoom from 1x to 2x, the movement is direct, but if you change from 3x to 2x, the zoom changes to 1x first and than to 2x in order to reach the position from the same direction.
The LV (fully armed) Microscope Control Pad consists of the following portions:
The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.
Caution
This behavior is valid if the Limit Control option is turned off in LV Setup. If Limit Control is turned on, the warning appears if a move is ordered from a lens with magnification 5x (or less) to a lens with Working Distance less then 1 mm.
Press this button to move the Z drive to the utmost position away from the sample. Use it when changing objectives or slides on the XY stage or during XY stage initialization. When the button is de-selected, the Z drive moves back to the most recent position.
The current Z position is displayed in the edit box. A particular value can be inserted as well and confirmed by pressing Enter to move the Z-drive.
When using Z drive of the ECLIPSE LV series, make sure that the bottommost position of the Z drive (called Lower Limit) is set to avoid collision between the specimen and the objective.
Click this button to set the current Z position as the Lower Limit.
Click this button to discard the Lower Limit setting.
Intensity of episcopic and diascopic illumination sources can be controlled from here. LED illuminators are also supported.
75% sets the aperture stop to 75% of the objective pupil. The slider with the edit box set a custom aperture size.
Controls the connected shutter.
There are 4 available positions in the changer. The first two have fixed filters for bright field (BF) and dark field (DF) microscopy. The other two are available for any filters.
Use the button to select the filters from the database of filters.
Defines the observation mode.
The L200/L300 Microscope Control Pad consists of the following portions:
The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.
A halogen lamp or the Nikon Intensilight illumination device can be attached to the EPI port of the microscope. Both illuminators can be controlled from here (see below). If two halogen lamps are attached, the illumination control works as the EPI/DIA switcher.
These two arrows control the EPI aperture stop equally to the buttons on the microscope body.
Note
It is not possible to save the aperture status to an optical configuration. The only way to adjust it is by these two buttons (or the buttons on the body).
You can select from where the lamps will be controlled:
from the Control Pad
from the microscope body
The MA200 Inverted Metallurgical Microscope Control Pad consists of the following portions:
The motorized objective changer can be controlled via this portion of the window. If a move from position 1 to position 5 is ordered a warning appears. This behavior shall prevent high magnification objectives from striking the XY stage.
Indicates the current state of the MA Zoom device. There are three predefined magnifications (1x, 1,5x, 2x).
A halogen lamp or the Nikon Intensilight illumination device can be attached to the microscope and controlled from here. If the halogen lamp is attached to the microscope, internal (manual) or external (via NIS-Elements) control can be selected on the back of the microscope body. LED illuminators are also supported.
This measuring microscope connected to NIS-Elements provides two logical devices:
Two types of nosepiece can be attached to the microscope, manual or intelligent.
See Motorized, Intelligent and Manual Devices.
After adding the MM microscope to the system via Device Manager, the Nosepiece logical device becomes available. Display its control panel by the View > Acquisition Controls > Nosepiece command. The type of the nosepiece must by defined via the hardware control pad in order to be recognized by NIS-Elements.
The Measuring Microscopes Z drive may be moved using the View > Acquisition Controls > XYZ Navigation control panel.
The NI-E Pad (fully armed) consists of the following portions:
Control the objective changer by clicking on the buttons.
Use the button to select objectives from the database of objectives. See Assigning Objectives to Nosepiece Positions.
An adjustable optical zoom is installed on the Rear microscope port. If the light path is switched to Rear a slider appears enabling to adjust the zoom factor.
Press this button to move the Z drive to the utmost position away from the sample. Use it when changing objectives or slides on the XY stage or during XY stage initialization. When the button is de-selected, the Z drive moves back to the most recent position.
Press this button to move the Z drive to the lowest position and move the XY stage outward from the microscope body. This function is designed for easy changing of samples. When the button is turned off, the XY stage moves back to the original position but the Z drive stays at its lowest. It is because the user would have to re-focus anyway after a new specimen was inserted.
Turns off the
button but does not move the Z Drive anywhere and lets you refocus manually.Select a port where the light will be directed.
Use the arrow buttons to move the Z drive by the specified distance. Or, enter the absolute distance to the Z edit box.
Control panels for controlling DIA and Intensilight (EPI) illumination appear if the corresponding devices are connected.
Control the available filter changers by clicking the buttons.
Use the button to select filters from the database of filters.
Control these aperture-like devices by moving the slider or entering numbers to the nearby fields.
Control the condenser changer by selecting a position from the pull-down menu.
Use the button to select condenser from the database of filters.
This button turns off some of the motorized microscope parts in order to minimize electromagnetic radiation of the device which might affect the specimen.
Check this item to use Ni recipe for capturing ND experiments. Define which port is used for the recipe.

A configuration window appears where it is possible to specify microscope options.
Limits to XY stage and Z drive movements can be specified. Insert maximum and minimum distances for each axis. Just the maximum/minimum of the Z drive can be specified depending on its type (whether it moves the stage or the nosepiece). See for further details.
Select speed for XY and Z motors.
Accuracy can be set to XYZ stages equipped with encoder.
This function guards the stage position in between stage movements so that the stage stays perfectly in its last position thus improving the overall experiment accuracy.
Speed of the filter turrets can be reduced by selecting the Slow option.
Specify what type of shutters are connected to the ports on the back of the microscope body (none, EPI, DIA, AUX).
This option enables you to hide or show the Sleep button on the Ni-E Pad.
Displays the Current Settings to Microscope Mode at the bottom of the Ni-E Pad. Select to which microscope mode will be the current settings assigned. Editable name of the selected microscope mode is displayed next to this pull down menu. You can enter arbitrary four characters as a name. If you press the Assign to Microscope button, current settings are assigned to defined microscope mode.
Saves the configuration settings into a .txt file using
and imports the dialog settings from a .txt file using . During the import, it is possible to select which parameters are imported using the check boxes.The Ni-U and Ni-L microscopes are simplified versions of the advanced Ni-E microscope. Please see Nikon Ni-E Microscope for the description.
Note
Supported models are: 18, 25, 754T, 800, 800N, 1000, 1270, 1270i, 1500
Depending on the type and setup of your SMZ, the control pad can contain different options.
Select the first objective on your nosepiece.
Click Objective Calibration).
and follow the calibration instructions (see:Switch to the next objective, go to Calibration > Objectives and it.
If your microscope has more than two objectives, repeat step 3.
An intelligent nosepiece which indicates the current light-path setting (Stereo/Mono).
Click the
button to move the Z Drive up by the distance specified in the rightmost edit box. Click it again to move the Z drive back to its original position.The โescaped or not escapedโ) to โnot escapedโ. E.g. If you pressed the sequence of buttons: - - , the Z drive would move twice the defined distance up.
button resets the escape status (This control enables you to specify the zoom value.
Objective magnification x Zoom = constant.
(Auto Link Zoom, supported by Nikon SMZ 25 only) can be activated to keep the magnification constant (same FOV) even if the objective lenses are switched:Zoom slider defines the zoom of the microscope (supported by SMZ 25 only). Use the edit box next to the slider to enter a precise zoom value.
assigns the current Zoom/Z setting to the position selected in the pull-down menu. To recall a setting, turn on and select the setting from the pull-down menu.
The Manual calibration window used for calibrating the current zoom magnification.
button opens theNote
For a correct zoom calibration on multiple SMZ objectives, follow this procedure:
Standard Z Drive control. The only difference from other microscopes is that the units used are millimeters.
Note
Because the minimum step of the SMZ Z drive is 1.6 ยตm, it is not possible to move exactly by the amounts displayed in the XYZ Navigation (View > Acquisition Controls > XYZ Navigation ) and by values lower than this step. XYZ navigation is a generic GUI and its values are not changing via Z drive, so SMZ moves to the next step with the best precision which it can. Use 1 and 0.5 up arrow buttons to get over your desired Z value and then precise the number by 0.05 and 0.003 down arrows.
Control panels for controlling DIA and Intensilight (EPI) illumination appear if the corresponding devices are connected.
The microscope may be equipped with a motorized (or intelligent) filter turret.

A configuration window appears where it is possible to specify microscope options.
The Nikon SMZ product line is very variable, some models are fully manual, some contains intelligent or motorized parts. If you plan to use your SMZ in the manual mode, connect the microscope as Nikon SMZ Manual in the device manager.
See Motorized, Intelligent and Manual Devices.
Pad Options
Specify the objectives mounted to your nosepiece via the button.
Zoom factors selected for display in the configuration window.
You can modify calibration for the selected Zoom. Click the button and proceed to manual calibration.
If a turret is selected in the configuration window, you can select the used filters via the button.
Click this button to specify properties of the SMZ microscope used.
Microscope model. Other options in this window depends on the selection.
Select how many objective positions will be displayed in the pad.
Is the turret mounted? Select this option.
Having this accessory mounted on the microscope changes the zoom factor. Select it if you have it.
You can modify the list of displayed zoom buttons by the
and buttons. The button loads the default list of zooms and their calibrations.The Ti2 (fully armed) Microscope Control Pad consists of the following portions:
Control the objective changer by clicking on the buttons.
Use the button to select objectives from the database of objectives. See Assigning Objectives to Nosepiece Positions.
Note
If the installed nosepiece supports the correction collar, the position number 5 gets disabled.
Press this button to move the Z drive to the utmost position away from the sample. Use it when changing objectives or slides on the XY stage or during XY stage initialization. When the button is de-selected, the Z drive moves back to the most recent position.
The
button is displayed in the escape mode and can be used to clear the escape flag - i.e. stay at the current Z escape position and unlock the Z control.Camera / EYE notifies about the light path direction if an eyepiece having two paths (eyepiece and camera) is connected.
The light path control buttons correspond to the buttons placed on the microscope body.
Note
Bertrand lens is automatically removed from the light path when switching from EYE to Camera.
Movements of the integrated Z drive are performed by the up/down arrows. Accuracy can be set in the pull-down menu on the right. The Accuracy setting affects the Z drive speed slightly.
This section controls the objective correction collar. The slider and the edit box defines the correction collar position.
Note
If the correction collar is used with a confocal microscope, the recommended pinhole size is 1.2 AU.

Click this button to open the ACC Calibration dialog window. See Automatic Correction Collar Adjustment.
Use the slider to set the correction collar value. Hover the cursor over the slider to display a tooltip table with the cover glass thickness and recommended correction collar values for particular temperatures.

Named presets of the correction collar values to the pull-down menu on the left of this button. Here you can add values to the list, name them and sort them.
If the dispenser device is installed correctly and a water immersion objective is activated in the microscope pad, these three buttons can be used:
Starts supplying water to the tubes and on the objective.
Water evaporates in time. Once initiated, automatic replenishment is started and the period is set to 30 minutes. Shall the water be replenished earlier, use this button.
Empties the tubing. Use this when you have finished the experiment.
See also Using Water Immersion Dispenser.
This optional accessory can be controlled via the pad.
The main
button turns the perfect focus system on and applies the current PFS offset. If the current objective supports two materials of the sample carrier (glass slide, plastic dish), a pull-down menu appears next to this button where the actual material in the sample holder shall be selected.This button moves the motorized dichroic mirror for the PFS in and out. PFS works only if set to the In position.
If the PFS is on, the offset value can be typed right in the edit box.

Turns the PFS on and finds the offset value so that it corresponds to the current Z-drive position.
The light source can be turned on/off using the button. Use the slider to set the light intensity or enter an exact value (percents or voltage) into the edit box.
In case of LED DIA Illumination the Optical Configurations as an active shutter.
button is added to the top toolbar so that it can be used inAll available Ti2 shutters are controlled from here. When Nikon Intensilight is connected,
shutter appears. FL-Up and FL-Lo shutters correspond with the shutters in the filter turret.Control the available filter changers by clicking the buttons.
Use the button to select filters from the database of filters.
Control the condenser changer by selecting a position from the pull-down menu.
Use the button to select condenser from the database of filters.
Can be used to select the External Phase Ring for the eyepiece.
This mode is used to prevent any potentially harmful effects caused by the electronics inside the microscope which may be potentially harmful to the specimen or otherwise interfere with the acquisition process. Use the button to temporarily turn the electronics off and back on.
Display or hide this option in the General tab of the microscope configuration dialog.

Opens the configuration window.
Note
If a halogen lamp is connected to the system, โMount Halogen Lampโ option has to be checked. This lamp requires an amount of time to be fully turned on and off. You can adjust the On/Off Time [msec] in the edit boxes which delays all experiments using the halogen lamp on/off command.
Does as it says.
Turns the front panel LED diode ON.
โAllowTi2ParfocalCorrectionโ has to be enabled first in the SetupOptions.xml
file in the NIS-Elements installation folder.
Set the ratio for dividing light between ports when the Right port is selected.
Saves the configuration settings into a .txt file using
and imports the dialog settings from a .txt file using . During the import, it is possible to select which parameters are imported using the check boxes.Sets the software limits for the XY, speed and accuracy.
Sets the software limit, speed, Open/Closed Loop Control, accuracy and Keep Z position.
The main box of the Ti2-E has two connectors which can be used to establish a connection using 8 BNC I/O with additional 16 BNC with the Extension box. Select the usage of the BNC connectors in the pull-down menus:
Enables the NIDAQ resource and controls accessories in the Triggered Acquisition and Illumination Sequence windows. PFI is selected per camera.
FIRE signal from the camera is connected to this pin. It enables the user to use the Ti Recipe (recipe is turned on).
copies the camera out signal (works like a hardware โT connectorโ).
Output signal: Hi = something is moving, Lo = nothing is moving.
Hi = in focus, Lo = other states
Hi = in range but not in focus, Lo = other states
Hi = in focus or in range and over the sample, Lo = other states
Hi = in range, Lo = out of range
If set, Output Signal Toggle on extended IO %d (two state button) or Output Signal High on extended IO %d Firmware Function in the Programmable Buttons tab can be assigned to a joystick button or microscope body button.
If set, Output Signal 5V 10ms on IO Firmware Function in the Programmable Buttons tab can be assigned to a joystick button or microscope body button.
Select this option to send the signal to the โtrigger inโ connector of a camera during a continuous scan. A continuous scan is used for example for fast detection of well plates used by the Sample Navigation panel.
Defines the programmable buttons on the microscope body and joystick. Each button can be used to run a specific macro (Run Macro), to run a command (Command List), to capture (Capture button), to switch to a selected light path (Select Lightpath) or to run a Firmware Function.
Note
When NIS-Elements is running, do not assign any Firmware Functions from outside tools and applications. If you still need to assign functions from external tools and applications, close NIS-Elements first.
Macro commands work with the programmable buttons only during the NIS-Elements session and override the firmware functions. When NIS-Elements is closed, all programmable button assignments are restored to the firmware state.
Mounts the LAPP system. If a motorized or intelligent branch is mounted, it is automatically added and cannot be removed.
See: View > Acquisition Controls > LAPP Pad, Motorized, Intelligent and Manual Devices.
Two motorized filter turrets can be installed on the microscope. Their position (Upper Layer or Lower Layer) is detected automatically. In some specific cases, the positions could end up switched. In this tab, you can select the position of each turret manually.
This function guards the stage position in between stage movements so that the stage stays perfectly in its last position thus improving the overall experiment accuracy.
The function cannot be used and is automatically turned off when using PFS.
Note
All the commands used in Ti pad for moving Ti accessories are recorded in the Command History (Macro > Command History).
All the microscope settings done inside NIS-Elements or in the Ti2Control App are stored in the microscope from which they are loaded on startup.
Runs the Ti2 Control where it is possible to set microscope accessories and options and execute โAssist Guideโ suitable for making preparations of microscope observations. See the โTi2 Control manualโ book for more information.
Place the fluorescent beads sample (sample where PSF can be obtained) onto the stage. Or else, place the beads sample or biological specimen on the stage. In case of beads, find the less densely populated region where the number of beads within the FOV is about 10.
Click on the
Live button and enable in the Live window.
Display the Histogram (View > Visualization Controls > Histogram
) and adjust light and exposure time so that no saturated pixels are present in the image. After the adjustment, the live image.
Click on the
button in Ti2 Pad to show the ACC Calibration dialog window.
Select the Sample Type in this window. Click on to show the Live window.
Move the motorized stage to place the target image (isolated beads or cells of interest) in the center of the FOV and bring it into focus. Set ROI to the target image in the Live window.
Click on ACC Calibration window to close the Live window.
in theSample Z Range and Search Range should be set as follows based on the type of objective.
Table 2.
Objective/Sample Type Sample Z Range Search Range Range* Z Step* Min Max Step Count 100x Oil, Beads 5.0 0.20 -4960 13280 7 100x Oil, Biological Specimen 2.5 0.05 -4960 13280 7 60x Water, Beads 30.0 0.20 -6240 6240 7 60x Water, Biological Specimen 15.0 0.05 -6240 6240 7 Click ACC Calibration window to run auto correction. Z-Stack image acquisition is run and correction collar is adjusted automatically. When the adjustment is finished successfully, the correction collar is moved to the adjustment position.
in the
Note
Approximately a two hour warm-up period is required after switching the microscope on.
For the correction ring adjustment, images acquired by the Epi-FL illuminator unit can be used as well as N-SIM images.
Hardware restrictions
Please use one of the following cameras: Andor iXon3, Andor iXon Ultra or Hamamatsu ORCA Flash4.0. For the Z-Stage, please use MCL Piezo Z. If ACC is performed with a camera or a Z-stage different than those mentioned, correct results cannot be obtained.
Place the fluorescent beads sample (sample where PSF can be obtained) onto the stage.
Find the less densely populated region where the number of beads within the FOV is about 10.
Click on the
Live button and enable in the Live window.
Display the Histogram (View > Visualization Controls > Histogram
) and adjust light and exposure time so that no saturated pixels are present in the image. After the adjustment, the live image.
Move the motorized stage to put the isolated fluorescent bead image in the center of the FOV and bring it into focus.
Tip
Vertical movement can be adjusted by the mouse wheel after clicking on the live display.
Click on the Z tab. Press and then to set the current focus position to the center of the stack.
Set the Z stack so that the Z-Step is 0.12~0.16 ฮผm with 25~30 Steps.
Click
to start image acquisition.On the acquired image window, click on the Show Slices View icon to show the cross-sectional images of XZ and YZ.
From the Mode pull-down menu select Slices.
Set the display magnification to 200%.
Open View > Visualization Controls > LUTs
and lower the upper limit of LUTs to make it easier to see the defocus spread.
Move the correction ring by one position to refocus.
Press ND Acquisition (About ND Acquisition) to set it to the center position of the Z stack.
of the Z series tab in theClick
to acquire Z stack images in the same way.When the capturing is done, move the position of the correction ring and repeat the same process.
By changing the position of the correction ring, compare the series of Z stacks and select the position so that the symmetry in the Z direction is the best. Adjustment of the correction ring is completed.
In the examples below, the symmetry is best at 0.18.
Note
Approximately a two hour warm-up period is required after switching the microscope on.
For the correction ring adjustment, images acquired by the Epi-FL illuminator unit can be used as well as N-SIM images.
Suppose the device is connected properly, a Water Immersion Dispenser section appears int the microscope pad. An immersion objective must be selected in the nosepiece to have all the buttons enabled.
Click the
button to add the first water on the objective. A progress dialog appears.Note
If the objective is in focus position, it will move to the escape position for the time needed for initialization.
Information about the time to the next automatic replenishment starts appearing below the buttons. E.g.: Next Replenishment in 30 min. If you need to replenish the water sooner than indicated, click the button.
After the experiment is finished, discharge water from the tubes by the
. This is important in order to keep the tubing clean. If you forget, the software will prompt you to discharge the dispenser upon exit.At last, do not forget to dry up the objective and the slide so that these parts stay clean too.
Tip
Behavior of the device can be customized in Windows registry on the following path:
HKEY_LOCAL_MACHINE\SOFTWARE\Laboratory Imaging\Misc\Ti2\WID
Settings for each objective can be edited as well (there are sub-folders named with objective codes).
Note
All Water Immersion Dispenser events are logged per microscope connection in a separate file into the following location:
C:\ProgramData\Laboratory Imaging\Platform\Logfiles\WIDRecord_*.log
where * is the date.
Most features of the Ti2-A Microscope are similar to the Ti2-E Microscope. Main difference is that Ti2-A microscope uses only intelligent models of Shutters, Filters and Condenser.
See Motorized, Intelligent and Manual Devices.
Please see Nikon Ti2-E Microscope for full description of the microscope features.