Objectives

If you want to perform measurements on captured images it is needed to calibrate the objective used to capture images. Whenever an image is captured through a calibrated objective, the image inherits its calibration.

Managing Objectives

Run the Calibration > Objectives command. The Objectives window appears:

A list of used objectives can be created within the Objectives window. For each objective, the objective name, position in the changer, storage status and the calibration are displayed. A Calculator icon next to the calibration value indicates that the calibration has been calculated from objective properties. If the calculator icon is missing, the calibration has been performed manually. Use the buttons on the right side to manage the objectives. Edit the properties of the selected objective in the bottom part of the window.

See Calibration > Objectives.

Assigning Objectives to Nosepiece Positions

  1. Display the microscope control pad - View > Acquisition Controls > Microscope Pad

  2. Click the setup button in the nosepiece section.

  3. A window appears which enables you to select one of the available objectives to the corresponding position.

  4. Each objective has its specifications displayed in the table. These specifications are not editable.

Objective Calibration

After you press the Recalibrate button in the Objectives window or when creating a new custom objective, the following window appears:

Select one of the calibration methods:

  • The Manual calibration lets you draw a distance into a picture and assign real length to it. The live image starts automatically and a window appears. Proceed to: Manual Calibration

  • If a motorized XY stage is available, the Auto and 4 points buttons appear. Proceed to: Objective Calibration Using XY stage.

Press OK to continue.

Superresolution Calibration 
This type of calibration is available after successful manual or automatic calibration. It triples the camera's resolution and uses fine stage movements to improve the resulting objective calibration. Click the Run SR Calibration button and wait till the objective is calibrated.

Manual Calibration

Manual calibration requires the user to specify a distance in the image. It can be either a measurable distance or the size of one pixel. The following procedure can be used for different purposes:

  1. The following window appears:

    1. If you know the size of one pixel, click the Pixel size... button.

      Enter the value and finish the calibration by the OK button.

    2. If you do not know the pixel size, you will need to specify distance in the image. For this purpose, use an image of a calibration slide or a ruler captured with your current system configuration.

  2. The distance is defined by placing lines to the image. If you are sure the calibration pattern is perpendicular to image edges, it is recommended to select either vertical or horizontal lines. Otherwise, select the parallel lines .

  3. Click into the image to place the first line and then the second line to specify the distance:

    ,

    You can modify the line position while holding the primary mouse button. After you release the button no further changes can be made. Place the second line in the intended position by another click.

    Draw the first line by clicking twice inside the image. The line can be moved and adjusted arbitrary by mouse. When satisfied, finish the first line by right-click. The second line can be placed by another click to the image, this time only to adjust the distance from the first line. The process is completed by right-click.

    Zoom on Click zooms the image after the mouse button is clicked so that the position of the point can be set precisely.

  4. Then the following dialog box appears:

  5. Enter the distance between the two lines and select correct units.

  6. Click OK to finish the calibration.

Objective Calibration Using XY stage

This procedure requires a motorized stage. Select one of the following methods:

Auto

The Auto method is fully automatic. Calibration is performed on a part of the live image marked by red square that is shown before the calibration is launched. You can also choose the channel to be used by auto-calibration.

NIS-Elements moves the motorized stage, acquires two images, and calculates the calibration from the shift of the images. The auto-calibration performs an estimation of the real relay lens zoom factor. If the estimation does not match the value specified in the Device Manager window, a warning appears.

It is recommended to cancel the auto calibration (click No), check the real zoom factor of your relay lens and set correct relay lens settings in the Device Manager.

Note

The success of this method depends on the texture of the specimen, its contrast, illumination, etc. If the combination of these factors is unsuitable, the auto calibration may fail. If it fails, try the following:

  • Move the stage to another area of the specimen to get a better texture.

  • Improve contrast by setting LUTs.

  • Check focus, re-focus if needed.

  • Turn on the shading correction (see Shading Correction).

4 points

If you select the 4 points method, the system draws four points on the screen (subsequently) and asks user to move one significant part of the specimen to each position. After all four steps are completed, the calibration is calculated from the moves of the stage.

Monitor Calibration

A real zoom factor can be displayed within the image window. In order for the value to be accurate, both the image calibration and the monitor calibration must be accurate. To calibrate the monitor:

  1. Display a context menu over a calibrated image and select the Show Physical Zoom command. The computed value appears in the top left corner of the image. If the current image is not calibrated, the box will show N/A.

  2. Right-click the box with the zoom value and select Monitor Calibration, the following window appears.

    Caution

    The dialog displays calibration of the monitor on which it is displayed. If you have multiple monitors, move the dialog between them.

  3. Check the Width and Height values and whether they match the real size of your monitor.

    Tip

    The detected monitor type is displayed in the window. If the information about the model is precise, it is likely that the automatic calibration is correct, otherwise, we recommend to switch to Custom monitor calibration.

  4. If you are not sure whether the detected calibration is correct, select Custom monitor calibration. Take a ruler or a measuring tape and measure width and height of your monitor. Enter these values and confirm the calibration by the OK button.

Adding Objectives to Objective Database

  1. Create an INI file containing specification of the new objectives (see the example INI below).

  2. Copy the INI file to NIS-Elements installation folder (where the executable file is placed).

  3. Start NIS-Elements. The objectives are loaded to the database automatically.

Example INI file: objectives.ini